OCETA Environmental Technology Profile
![[Cetac symbol]](../../images/cetac-i.gif)
The ELAN® 6000 is a compact, self-contained Inductively Coupled Plasma-Mass Spectrometry (ICP-MS) system, that combines detection sensitivity equal or superior to GFAA (graphite furnace atomic absorption) with the rapid multi-element sample throughput of ICP-OES (ICP optical emission spectroscopy). ICP-MS uses a high-temperature argon plasma to generate metal and elemental ions which are introduced into a mass spectrometer for separation and quantitation. The ELAN 6000 can assess a sample, covering the full range of the elements in the periodic table, in less than two minutes. Figure 1 shows the ELAN 6000 ICP-Mass Spectrometer.

Figure 1: The ELAN 6000 in operation
Unique features of the ELAN 6000 include:
The ELAN 6000 is applied for the characterization of both inorganic and organic materials in the routine analysis of trace elements in drinking water, groundwater, surface water, wastewater, solid waste, industrial effluent, soil, sediment and atmospheric samples. All 12 primary contaminants, including Hg, currently regulated by the US Safe Drinking Water Act, can be determined by a single method, replacing the previous requirement of three or more instruments to achieve the same analysis.
Table 1 shows the ELAN 6000's instrument detection limits (IDLs) measured with US EPA Method 6020 and the recoveries of selected elements in estuarine sediment. The certified values were recovered within a few percent of the true value and a 15 ppb (ug/L) spike was recovered well within the 75-125% recovery limits.
Element |
ELAN IDL (ppb) |
Sample Conc. (mg/kg) |
% Rec. of Cert. Value |
% Rec. of Spike |
|---|---|---|---|---|
| Al | 0.04 | 548 | 84.2 | NA |
| Sb | 0.003 | 0.004 | 98.9 | 103 |
| As | 0.06 | 0.11 | 112 | 98.4 |
| Ba | 0.01 | ND | - | 98.9 |
| Be | 0.02 | 0.02 | - | 117 |
| Cd | 0.002 | 0.0004 | 104 | 93.6 |
| Cr | 0.1 | 0.78 | 97.4 | 97.0 |
| Co | 0.002 | 0.12 | 116 | 97.4 |
| Cu | 0.005 | 0.20 | 99.6 | 96.0 |
| Pb | 0.001 | 0.31 | 104 | 102 |
| Mn | 0.004 | 3.64 | 104 | 90.2 |
| Ni | 0.007 | 0.31 | 104 | 96.3 |
| Ag | 0.03 | ND | - | 94.9 |
| Tl | 0.0003 | ND | - | 99.6 |
| Zn | 0.015 | 1.52 | 109 | 98.2 |
Industries that currently use the ELAN ICP-MS technology include:

Figure 2: Schematic representation of the ELAN 6000
A schematic of the ELAN 6000 is shown in Figure 2. The sample is introduced into the plasma for destruction of the sample matrix and ionization. Ions pass from the torch region, at atmospheric pressure, to the quadrupole mass spectrometer, at vacuum pressures, through the interface region consisting of skimmer and sampling cones. The ion lens focuses the ions into the spectrometer, which separates the ions by mass-to-charge ratio and directs them to the detector where they are measured. The AutoLens ion optic system does on-the-fly computer-controlled lens optimization for maximum analyte ion signal with minimum interference.
ELAN NT software controls the hardware and methodology, including quality control checks, can correct for isobaric and molecular interferences, graphically displays data in real time, and provides quantitative analysis protocols for external calibration and standard additions. The ELAN 6000 was designed for unattended nighttime operation, thus increasing the productivity of the routine environmental laboratory. Table 2 lists some specifications for the ELAN 6000.
| Sensitivity: | >40 M cps/mg/L 115In |
|---|---|
| Long-Term Stability: | <4% RSC over 4 hours |
| Regulatory Compliance: | CSA, FCC, CE, and other standards |
| Power: | 200-240 V, 50/60 Hz |
| Size (without computer): | 72 cm x 99 cm x 110 cm |
| Weight: | 295 Kg |
The ELAN ICP-MS technology originated in 1983 at the University of Toronto with Dr. D. Douglas and Prof. J.B. French, working under contract to the Sciex Division of MDS Health Group. Sciex continued development and sales until 1986, when a joint venture was formed between Sciex and the Perkin-Elmer Corporation. Sciex develops and manufactures the ICP-MS at the Sciex facility in Toronto. Perkin Elmer provides a worldwide sales and service network. Over 600 ELAN ICP-MS Systems are installed worldwide, with over 40% used for environmental applications. PE-Sciex also offers the tandem quadrupole API-300 MS-MS and the single quadrupole API-100 MS with the LC ion spray interface. Perkin Elmer offers a comprehensive range of analytical instrumentation for the environmental monitoring of organics and inorganics.
Profile Print Date: 06/96
Catalogue # 07-003-H